Labtro Biological Safety Cabinet Class II A2 LT-BSC207 draws room air inward through the front opening to create an air barrier that helps protect the operator from aerosols and contaminants generated inside the work zone. The incoming air mixes with internally recirculated air and is passed through the U15 ULPA downflow filter, which supplies clean vertical laminar airflow over the work surface to protect the samples. Approximately 70% of the filtered air is recirculated back through the cabinet, while the remaining 30% is discharged through a separate ULPA exhaust filter to protect the surrounding environment. An EC brushless motor with closed-loop airflow control continuously adjusts fan output to compensate for increasing filter resistance and maintain the specified inflow and downflow velocities. Integrated sensors monitor airflow, filter pressure, sash position, and fan performance, while the touchscreen displays operating conditions and activates audible and visual alarms if parameters move outside safe limits. After operation, the programmable 254 nm UV lamp can be used for work-zone decontamination when the cabinet is not in active use.
Technical Specifications
| Classification | Class II Type A2 | Protection Type | Personnel, Product, and Environmental Protection |
|---|---|---|---|
| Airflow Pattern | 70% Recirculation / 30% Exhaust | Airflow Direction | Vertical Laminar Downflow |
| Exhaust Direction | Top Exhaust | Average Inflow Velocity | 0.50 m/s |
| Average Downflow Velocity | 0.33 m/s | Airflow Distribution | Full-area homogeneous vertical downflow |
| Airflow Regulation | Closed-loop Constant-Air-Volume Regulation | Airflow Compensation | Automatic compensation for filter loading |
| Airflow Monitoring | Continuous inflow and downflow velocity monitoring | Velocity Sensors | Integrated |
| Velocity Meter Accuracy | 0.001 m/s | Blower / Motor | EC Brushless DC Motor |
| Downflow Filter | Independent ULPA Filter | Exhaust Filter | Independent ULPA Filter |
| Filtration Configuration | Dual ULPA Filtration – Downflow and Exhaust | ULPA Classification | U15 |
| Filtration Efficiency | >99.999% at 0.1 to 0.3 µm | Controller | Microprocessor Control System |
| Display | 7-inch Color Capacitive Touchscreen | Displayed Airflow Parameters | Inflow Velocity and Downflow Velocity |
| Filter Monitoring | Filter Differential Pressure and Filter-Life Monitoring | Fan Monitoring | Fan Operating Status |
| Sash Monitoring | Front Sash Position / Status | Running-Time Display | Supported |
| UV Timer Display | Supported | System Information Display | Supported |
| Alarm Type | Audible and Visual | Inflow Alarm | Inflow Velocity Outside Preset Limit |
| Downflow Alarm | Downflow Velocity Outside Preset Limit | Fan Alarm | Fan / Blower Failure |
| Sash Alarm | Sash Position Outside Safe Operating Range | Filter Alarm | Filter Blockage / High Resistance |
| Filter-Life Alarm | Filter Lifetime / Maintenance Reminder | Work Chamber Material | 1.5 mm SUS304 / AISI 304 Stainless Steel |
| Work Chamber Design | One-piece formed construction with large-radius coved corners | Worktop Material | SUS304 Stainless Steel |
| Worktop Type | Removable without tools | Worktop Cleaning Access | Allows access to work surface and lower plenum for cleaning/decontamination |
| Side Panels | Tempered Safety Glass | Front Sash Material | Laminated Shatterproof Safety Glass |
| Front Sash Design | 10° Backward-Slanted Ergonomic Sash | Working Sash Opening | 180 mm |
| Maximum Sash Opening | 530 mm | Armrest | Integrated Ergonomic Armrest |
| Main Body Material | 1.2 mm Electro-Galvanized Steel | Surface Treatment | White Oven-Baked Electrostatic Powder-Coated Finish |
| Chemical Resistance | Resistant to Standard Laboratory Disinfectants | Illumination Type | LED |
| Standard Illumination | ≥700 lux | Enhanced Illumination | ≥900 lux configuration available |
| Lighting Design | Low-heat, shadow-minimized illumination | UV Lamp | 254 nm Germicidal UV |
| UV Lamp Power | 30 W | UV Timer | Programmable |
| UV Safety Interlock | Interlocked with Front Sash and Fan | Internal Electrical Sockets | 2 Waterproof / Drip-Proof Sockets |
| Vibration Level | ≤3 µm RMS | Base Stand | Height-Adjustable Support Frame |
| Power Supply | AC 220 V ±10%, 50/60 Hz | Electrical Requirement | Dedicated Grounded Supply |
| Maximum Electrical Load | 1500 VA | Internal Work Area Dimensions | 1300 × 672 × 630 mm |
| External Dimensions with Base Stand | 1480 × 790 × 2080 mm | Net Weight | 270 kg |
| Packing Dimensions | 1600 × 1020 × 1740 mm | Gross Weight | 302 kg |
Features & Applications
Key Features
Class II Type A2 protection system
↳ Provides simultaneous personnel, product, and environmental protection
70% recirculation and 30% exhaust
↳ Maintains controlled containment while conserving conditioned laboratory air
Dual U15 ULPA filtration
↳ Independent downflow and exhaust filters provide >99.999% efficiency at 0.1 to 0.3 µm
Closed-loop airflow regulation
↳ EC brushless motor automatically compensates for filter loading to maintain stable airflow
Continuous airflow monitoring
↳ Tracks inflow and downflow velocities with integrated sensors
7-inch color touchscreen
↳ Displays airflow velocity, filter differential pressure, filter life, fan status, sash condition, timers, and alarms
Comprehensive alarm system
↳ Provides audible and visual warnings for airflow deviation, fan failure, sash position, filter blockage, and maintenance requirements
Removable SUS304 worktop
↳ Enables thorough cleaning and decontamination of the work surface and lower plenum
One-piece stainless-steel work chamber
↳ Large-radius corners minimize difficult-to-clean joints and support routine decontamination
10° ergonomic safety-glass sash
↳ Provides comfortable access and clear visibility during operation
Tempered-glass side panels
↳ Improve lateral visibility of the work zone
Programmable 254 nm UV sterilization
↳ Supports work-zone decontamination with sash and fan safety interlocks
Dual waterproof electrical outlets
↳ Allow operation of compatible small laboratory equipment inside the workspace
Height-adjustable support stand
↳ Provides stable and ergonomic cabinet positioning
Applications
Clinical Laboratories → Processing potentially infectious diagnostic and patient samples
Cell Culture Laboratories → Aseptic handling of cells, tissue cultures, and related biological materials
Biotechnology Laboratories → DNA/RNA procedures, hybridization, media preparation, and biological research workflows
Pharmaceutical Laboratories → Microbiological quality control, sterile preparation, and biological sample handling
Hospital Laboratories → Controlled preparation and manipulation of clinical biological specimens
Biomedical Research Laboratories → Procedures requiring personnel, sample, and environmental containment
Quality Control Laboratories → Controlled microbiological testing and aseptic sample processing
Product Details
Biological Safety Cabinet Class II A2 LT-BSC207
Standard Accessories:
- Height-Adjustable Base Stand
- Downflow U15 ULPA Filter
- Exhaust U15 ULPA Filter
- 254 nm UV Germicidal Lamp
- LED Illumination Lamp
- 2 Waterproof Internal Electrical Sockets
- 7-inch Touchscreen Controller
- Airflow Velocity Sensors
- Filter Differential-Pressure Monitoring System
- Integrated Armrest
- Power Cord / Electrical Connection
Optional Accessories:
Complete ULPA Filter Set
Spare UV Lamp
Spare LED Lamp