Overview
Labtro Ductless Fume Hood LT-DLFH102 draws contaminated laboratory air inward through the front opening and directs it through a multi-stage filtration system. Larger particles are first captured by the pre-filter, while fine particulates are retained by the HEPA filter and chemical vapors are adsorbed by the activated-carbon filter. After filtration, the treated air is recirculated back into the laboratory, eliminating the need for an external exhaust duct. The VAV airflow control system automatically regulates air movement to maintain suitable operating airflow, with an air-handling capacity of 220 to 1200 m³/h. The LCD touchscreen displays air speed, operating time, and filter condition, while the saturation-monitoring system triggers audible and visual alarms at approximately 80% filter saturation. LED lighting supports clear visibility inside the work area, while UV sterilization can be used for internal decontamination when required. The system therefore controls exposure to compatible laboratory fumes by continuously capturing, filtering, monitoring, and recirculating the treated air.
Technical Specifications
| System Configuration | Ductless / Recirculating | Airflow Control | VAV Automatic Airflow Regulation |
|---|---|---|---|
| Air Handling Capacity | 220 to 1200 m³/h | Airflow Adjustment | 3 Levels |
| Display | LCD Touchscreen | Air-Speed Display | Supported |
| Operating-Time Display | Supported | Filtration System | Multi-stage Modular Filtration |
| Primary Filter | Pre-filter | Particulate Filter | HEPA Filter |
| Chemical Filter | Activated Carbon Filter | Filtration Efficiency | Up to 99.9% for filtration configuration |
| Filter Saturation Monitoring | Real-time LCD Display | Filter Saturation Alarm Level | Audible and visual alert at 80% |
| Alarm Type | Audible and Visual | Fan Power | 138 W ±15% |
| Rated Power | 2000 W | Noise Level | Approximately 50 dB |
| Cabinet Material | 1.0 mm Cold-Rolled Steel | Surface Treatment | Epoxy Resin Powder Coating |
| Chemical Resistance | Acid- and Alkali-Resistant Construction | Work Surface | 12.7 mm Solid Physicochemical Board |
| Alternative Worktops | Ceramic, Marble or Stainless Steel configuration available | Side Panels | 8 mm Acrylic Transparent Panels |
| Front Window | Foldable Transparent Front Window | LED Lighting | 10 W LED Lamp ×1 |
| UV Lamp | 20 W UV Germicidal Lamp | Internal Socket | 10 A Socket |
| Electrical Protection | Built-in Current Overload Protection | Mobility | Universal Casters |
| Power Supply | AC 220 V, 50 Hz | Optional Voltage | 380 V configuration |
| External Dimensions | 1200 × 650 × 1900 mm | Internal Working Space | 1120 × 600 × 700 mm |
| Filtration Technology | Vertically arranged modular activated-carbon and HEPA filtration | Packing Dimensions | 1300 × 880 × 1200 mm |
| Gross Weight | 220 kg |
Features & Applications
Key Features
Ductless recirculating design
↳ Eliminates the need for connection to a conventional external exhaust duct
220 to 1200 m³/h air handling
↳ Supports controlled capture and treatment of laboratory fumes
Multi-stage filtration
↳ Combines pre-filter, HEPA and activated-carbon filtration
Modular chemical filtration
↳ Allows activated-carbon and HEPA filters to be configured according to laboratory requirements
Automatic VAV airflow regulation
↳ Adjusts airflow during operating conditions
LCD touchscreen
↳ Provides convenient airflow, filter and operating-time monitoring
Filter saturation monitoring
↳ Displays filter condition continuously during operation
Audible and visual filter alarm
↳ Provides warning when filter saturation reaches 80%
Chemical-resistant worktop
↳ 12.7 mm solid physicochemical board supports routine chemical handling
Corrosion-resistant construction
↳ Epoxy-coated cold-rolled steel provides resistance against acids and alkalis
Current overload protection
↳ Protects the electrical system during abnormal operating conditions
Mobile integral construction
↳ Universal casters support convenient positioning within the laboratory
Applications
Scientific Research Institutes → Containment of chemical vapors during experimental work
Disease Control & Quarantine Laboratories → Sample and reagent procedures involving chemical fumes
Environmental Laboratories → Handling reagents and chemicals during environmental testing
Life-Science Laboratories → Chemical preparation and analytical procedures
Pharmaceutical Manufacturing → Chemical preparation and quality-control workflows
Food Laboratories → Reagent preparation and analytical testing
Chemical Industries → Laboratory-scale chemical handling and testing
Feed & Flavor Industries → Chemical and formulation quality-control operations
Product Details