Pulse Vacuum Autoclave LT-PVA401

Pulse Vacuum Autoclave LT-PVA401

Pulse Vacuum Autoclave LT-PVA401 features a 150 L chamber capacity and operates up to 150℃. Designed with a horizontal rectangular jacketed chamber. It supports heat retention and efficient drying through vacuum dehumidification at the end of each cycle. Equipped with PLC-based control, pneumatic door interlocking, and integrated safety monitoring for controlled operation. Our Pulse Vacuum Autoclave is widely used in CSSD, operating rooms, laboratories, and pharmaceutical facilities for sterilizing instruments, dressings, devices, and liquids.

Qty:

$ 16,830.00

Specifications

Capacity 150 L
Temperature 150 ℃
Working Temperature 105 to 138 ℃
Pressure 0.3 MPa
Pressure of Inner Chamber -0.1 to 0.3 MPa
Working Pressure 0.25 MPa
Vacuum -0.086 MPa
Limit Vacuum -97 kPa
Vacuum Leakage Rate ≤0.13KPa/min
Safety Valve 0.28 MPa
Temperature Control Precision 0 to 2 degree Celsius
Temperature Uniformity ± 1℃
Temperature Equilibrium <+2.0°C
Vacuum Pulsation Amplitude -80 to 0 kPa
Steam Consumption ≤14
Water Consumption ≤190
Vacuum Pulsation Times 1 to 99
Cross Pressure Pulsation Amplitude -80 to 0 kPa
Cross Pressure Pulsation Times 1 to 99
Positive Pressure Pulsation Times 1 to 3
Water Pressure 0.1 to 0.2 MPa
Compressed Air Pressure 0.4 to 0.7 MPa
External Steam Pressure 0.3 to 0.7 MPa
Power AC 380 V, 50 Hz( Three-phase five-wire )
Dimension 700 × 450 × 480 mm
Packaing dimension 950 × 1130 × 1760 mm
Net Weight 700 Kg
Gross weight 750 Kg

Features

  • Siemens PLC 
  • Touch-screen control
  • Jacketed stainless-steel chamber 
  • Pneumatic door sealing
  • Interlocking for safe operation
  • F0 and temperature-time sterilization monitoring
  • Multiple programs with data recording
  • Multi-level safety protection
  • Cycle recording
  • Safety alarms
  • Saturated steam 

Application

Pulse Vacuum Autoclave LT-PVA401 used in hospitals, laboratories, pharmaceutical and research facilities for sterilizing instruments, textiles, glassware, media, reusable medical equipment, biohazard waste treatment, and for CSSD processes.