Air Sampler
Air Sampler
Autoclave
Autoclave
Balance
Balance
Microscope
Microscope
Centrifuge
Centrifuge
Bath
Bath
Photometer
Photometer
View All Categories
Search suggestions will appear here.
Get Quote 0
Menu
Get Quote

Analytical HPLC System LT-AHS103

Product Code: LT-AHS103

Labtro Analytical HPLC System LT-AHS103 operates with 70 MPa maximum pressure keeps sample analysis moving from injection through compound separation and detector response in one continuous process. Organized solvent delivery maintains a steady analytical sequence while generated chromatograms support result verification and record keeping. Continuous sample processing improves laboratory throughput with clear, repeatable analytical data for routine testing.

Key Features
Binary high-pressure solvent delivery
Stable gradient flow control
Precise flow regulation
Continuous laboratory operation
Integrated chromatography communication
Modular HPLC configuration
$ 36,850.00

Qty:

Catalog
Need technical specifications, pricing or application assistance? Contact our product specialists.
Technical Data

Product Specifications

Pump Type
Binary high-pressure constant flow pump
Pump Flow Range
0.001 to 10.000 mL/min
Pump Flow Accuracy
≤ ±0.2%
Pump Flow Stability
RSD ≤ 0.06%
Pump Maximum Pressure
70 MPa
Pump Pressure Display Error
±0.3 MPa
Pump Pressure Pulsation
≤ 0.1 MPa
Pump Gradient Mixing Accuracy
≤ ±0.5%
Pump Gradient Composition Accuracy
≤ 0.1% RSD
Pump Gradient Composition Range
0 to 100%
Pump Communication
UDP
Pump Power Consumption
300 W
Pump Power Supply
AC 110–220 V, 50–60 Hz
Pump Dimension
450 × 350 × 180 mm
Pump Weight
18 kg
Autosampler Type
Automatic liquid sampler
Autosampler Injection Range
0–100 μL (Standard), 1–2000 μL (Optional)
Autosampler Injection Speed
≤10 s
Autosampler Sample Residue
<0.002%
Autosampler Injection Repeatability
RSD <0.2%
Autosampler Capacity
120 × 2 mL or 210 × 1 mL vials
Autosampler Maximum Pressure
45 MPa
Autosampler Power Consumption
30 W
Autosampler Power Supply
AC 110–220 V, 50–60 Hz
Autosampler Dimension
440 × 378 × 180 mm
Autosampler Weight
13 kg
UV Detector Type
UV-Visible detector
Detector Wavelength Range
190 to 800 nm
Detector Wavelength Accuracy
±1.0 nm
Detector Wavelength Repeatability
≤ ±0.1 nm
Detector Spectral Bandwidth
8 nm
Detector Noise
≤ ±0.25 × 10⁻⁵ AU
Drift
≤0.5 × 10⁻⁴ AU/h
Detector Linearity
≥2.5 AU
Minimum Detectable Concentration
1 × 10⁻⁹ g/mL
Light Source
Deuterium and Tungsten lamps
Detector Maximum Pressure
1500 psi
Detector Sampling Frequency
5–80 Hz
Detector Time Constant
0.1–5.0 s
Detector Power Consumption
100 W
Detector Power Supply
AC 110–220 V, 50–60 Hz
Detector Dimension
440 × 378 × 160 mm
Weight
10 kg
Diode Array Detector Type
Diode Array Detector (DAD)
Array Elements
512
Wavelength Range
190 to 800 nm
Spectral Resolution
2.4 nm
Array Resolution
1.2 nm
Wavelength Accuracy
±1.0 nm
Dynamic Noise
≤ ±1.0 × 10⁻⁵ AU
Baseline Drift
1.0 × 10⁻⁴ AU/h
Linearity
≥2.0 AU
Lamp
Deuterium lamp
Detection Pool
10 mm, 11 μL
Maximum Pressure
1000 psi
Power Supply
AC 110–220 V, 50–60 Hz
Column Oven Type
Column oven
Temperature Range
Ambient +5°C to 99°C
Temperature Accuracy
±0.1°C
Temperature Precision
≤0.1°C
Communication
RS485 / UDP
Power Consumption
110 W
Oven Power Supply
AC 220 V ±10%, 50 Hz
Dimension
120 × 65 × 570 mm
Oven Weight
4.2 kg
Performance

Features & Benefits

  • Binary high-pressure solvent delivery
  • Stable gradient flow control
  • Precise flow regulation
  • Continuous laboratory operation
  • Integrated chromatography communication
  • Modular HPLC configuration
Laboratory Uses

Applications

Solvent Delivery: Maintains stable mobile phase flow.

Gradient Elution: Mixes solvents for separation.

Sample Analysis: Supports consistent analytical runs.

Method Execution: Follows programmed flow conditions.

Column Operation: Delivers steady pressure to the column.