Biological Safety Cabinet Class II A2 LT-BSC202

Biological Safety Cabinet Class II A2 LT-BSC202 with 70% internal air recirculation and 30% exhaust for maintaining sterile working conditions. It features a microprocessor-based airflow control system that automatically adjusts airflow volume. Equipped with overload protection for stable and reliable operation.  Built-in large LCD screen and intuitive interface offers real-time monitoring. Our Biosafety Cabinet is ideal for sterile procedures involving pathogenic microorganisms or delicate biological materials.

Qty:

$ 4,043.00

Specifications

Type of BSC Class Ⅱ Type A2
Airflow Mode 70% Cycle, 30% Exhaust
Exhaust Direction Top Exhaust
Average Inflow Velocity 0.50m/s
Average Downflow Velocity 0.33m/s
Inflow Airflow Volume 446m3/h
Downflow Airflow Volume 1038m3/h
Exhaust Airflow Volume 446m3/h
Velocity Meter Accuracy 0.001m/s
Cleanliness ULPA: Class 3(ISO 14644.1); Class 100 (Fed 209E)
Filter Typical Efficiency Greater Than 99.999% at 0.1 to 0.3 Micron, ULPA as per IEST-RP-CC001.3 USA
Total Colony in Impaction Sampler (Operator Protection) ≤10CFU/Time
Total Colony in Slit Type Sampler (Operator Protection) ≤5CFU/Time
Total Colony in Culture Dish (Product Protection) ≤5CFU/Time
Total Colony in Culture Dish (Cross Contamination) ≤2CFU/Time
Personnel Protection Test KI-Discus Containment and Microbiological Testing are Performed
Blower AC
Downflow Filter ULPA
Exhaust Filter ULPA
Main Body 1.2mm (xxx'') 18 Gauge Electro-Galvanized Steel with White Oven-Baked Powder-Coated Finish
Work Zone 1.5mm (xxx'') 16 Gauge Stainless Steel, Type304, Single-Piece or 3-Piece
Side Walls 1.5mm (Xxx'') 16 Gauge Stainless Steel, Type304
Sash Glass Thickness 5mm
Sash Glass Type Tempered Glass, UV-Proof
Sash Window Working Opening 180mm
Sash Maximum Opening 530mm
Fluorescent Light Intensity ≥700lux
RMS ≤3um
Sound Emission (dBA) ≦65dBA
Dis, play 7 inches touch screen
UV Timer Yes
Main Power Switch Yes
Alarm Type Audible + Visual
Alarm Function Inflow Velocity, Downflow Velocity, Sash Height Limit, Filter Blocked, Filter Lifetime Remind
Cabinet Full Load Amps (FLA) 2.2A
Normal Size 4 Feet
Usable Working Area 0.8736m2, 9.40 sq.ft.
Stand Height 640mm (Adjustable)
Power Consumption 1500W
Power Supply AC220±10%,50/60H
Internal Work Area Dimensions 1300 × 672 × 630mm
External Dimensions with Base Stand 1480 × 790 × 2080mm
Packing Dimension 1610 × 950 × 2220mm
Net Weight 270 Kg
Gross Weight 300 Kg

Features

  • Waterproof interface
  • Diagnostic button
  • Strong UV light
  • Energy-efficient motor
  • Bright fluorescent lighting

Standard Accessories

  • Fluorescent lamp – 30W × 2

    Ultraviolet lamp - 30W × 1

    Receptacle

    Detachable Base Stand

Optional Accessories

  • Base Stand

Application

Biological Safety Cabinet Class II A2 LT-BSC202 is essential for sterile procedures involving pathogenic microorganisms or sensitive biological materials.

Expert Help Desk